COLLEAGUE COLOR PUMP CE FRENCHVERSION SINGLE
Report
- Report Number
- 6000001-2011-02314
- Event Type
- Malfunction
- Date Received
- March 29, 2011
- Date of Event
- January 17, 2011
- Report Date
- January 31, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Removal / Correction Number
- 6000001-12/1/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE DEVICE WAS EVALUATED ON-SITE AT THE CUSTOMER FACILITY. THE CONDITION OF DAMAGED BATTERY ALARM WAS CONFIRMED THROUGH THE EVENT HISTORY DUE TO DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND HARNESS WERE REPLACED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW UP MEDWATCH WILL BE SUBMITTED. THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT HAVE A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. (B)(4).
THERE IS AN ONGOING CAPA INVESTIGATION, MDQ-CAPA (B)(4) ASSOCIATED WITH THIS REPORT. (B)(4).
COLLEAGUE INFUSION PUMP WAS REPORTED ORIGINALLY FOR A DAMAGED BATTERY ALARM. IT IS UNKNOWN WHEN THE REPORTED CONDITION OCCURRED. NO PATIENT INJURY OR MEDICAL INTERVENTION OCCURRED. DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER (B)(4) PERSONNEL, THE DEVICE WAS FOUND TO HAVE EXPERIENCED A BATTERY DEPLETED SET ALARM WHICH INTERRUPTED DELIVERY. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.63.92 CATEGORIZED AS REMEDIATED.