XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Report
- Report Number
- 2024168-2011-02104
- Event Type
- Injury
- Date Received
- March 29, 2011
- Date of Event
- March 2, 2011
- Report Date
- March 4, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- NIQ
- PMA / PMN Number
- P070015
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THERE WAS NO REPORTED PRODUCT DEFICIENCY. DISSECTION IS A KNOWN ADVERSE EVENT AS LISTED IN THE XIENCE V INSTRUCTIONS FOR USE. ALTHOUGH A CONCLUSIVE CAUSE FOR THE REPORTED PATIENT EFFECTS AND THE RELATIONSHIP TO THE DEVICE, IF ANY, CANNOT BE DETERMINED, THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY WITH RESPECT TO MANUFACTURE, DESIGN, OR LABELING. ALL STENT DELIVERY SYSTEMS ARE SUBJECTED TO VISUAL INSPECTION. IN ADDITION, A QUALITY CONTROL AUDIT INSPECTION IS USED TO VERIFY THE PRODUCT QUALITY.
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE MID RIGHT CORONARY ARTERY WITH HEAVY TORTUOSITY AND HEAVY CALCIFICATION. DURING DEPLOYMENT OF THE 3.5 X 23 XIENCE V, A DISSECTION WAS OBSERVED, AND WAS TREATED WITH A 3.5 X 15 XIENCE V. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM | DRUG ELUTING CORONARY STENT SYSTEM | NIQ | AV-TEMECULA-CT | 0071942 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Required Intervention | GUIDE WIRE: BMW |