FDA Adverse Event Malfunction Summary report: N

CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V

MDR report key: 20325001 · Received September 27, 2024

Report

Report Number
2249723-2024-0003957
Event Type
Malfunction
Date Received
September 27, 2024
Date of Event
September 4, 2024
Report Date
March 11, 2025
Manufacturer
DATASCOPE CORP.
Product Code
DSP
UDI-DI
10607567107882
PMA / PMN Number
K063525
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 0

UPDATED FIELDS: B4, D8, D9(DEVICE AVAILABLE FOR EVAL, RETURN TO MANUFACTURE DATE), G1, G3, G6, H2, H3, H6(TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, COMPONENT CODES, INVESTIGATION CONCLUSIONS), H11. FSE REPLACED COMPRESSOR FAN (0119-00-0149). COMPLETED THE SCHEDULED PM. FSE NOTED ECG LEADS MISSING/NOT PRESENT, ADVISED THE INHOUSE BIOMED TEAM TO GARNISH NEW ECG CABLES A QUOTE WILL BE PROVIDED UPON REQUEST. RAN ALL THE TESTS IN ACCORDANCE TO THE MANUFACTURER'S SPECIFICATIONS. ANALYSIS AND TESTING DEPT. (FAT) WAYNE, NJ CF 10 DEC 2024. THE FAILURE ANALYSIS AND TESTING DEPT. RECEIVED PART NUMBER 0119-00-0149 FAN COMPRESSOR SERIAL NUMBER N/A WITH A REPORTED UNIT FAILURE OF THE FAN MAKING NOISE WHEN TURNED ON. THE FAILURE ANALYSIS AND TESTING DEPT. PERFORMED A VISUAL INSPECTION AND OBSERVED A BROKEN BLADE OF THE FAN. POSSIBLY, A PROBABLE CAUSE OF THE FAN BEING NOISY WHEN THE UNIT WAS TURNED ON IS, THE BLADE WAS IN THE PROCESS OF BEING SHEARED OFF WHEN THE UNIT WAS TURNED ON, OR A PIECE OF THE BLADE WAS HANGING AND PRODUCING THE NOISE. HOW THE FAN WAS DAMAGED IS IMPOSSIBLE TOO EXPLAIN. THE FAN IS PROTECTED BY A SIDE PANEL THAT COMPLETELY COVERS THE FAN. RETAINING THE FAN IN THE FAT DEPT. PER PROCEDURE NUMBER (B)(4).

Additional Manufacturer Narrative · 0

CORRECTED DATA: B5, D5, E1 (INITIAL REPORTER AND EMAIL), E2, E3, E4. UPDATED DATA: B4, G2, G3, G6, H2, H11.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN TURNED ON, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) FAN IS MAKING NOISE WHEN TURNED ON. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING PREVENTATIVE MAINTENANCE BY GETINGE FIELD SERVICE ENGINEER, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) FAN IS MAKING NOISE WHEN TURNED ON. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1433084 CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. 0998-00-3023-53 10607567107882

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown