FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM

MDR report key: 2032453 · Received March 25, 2011

Report

Report Number
2953200-2011-00715
Event Type
Injury
Date Received
March 25, 2011
Date of Event
February 25, 2011
Report Date
February 25, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS: (STENT GRAFT MIGRATION, ENDOLEAK). (SHORT DISTAL LANDING ZONE AT IMPLANT). CONCLUSION: (SHORT DISTAL LANDING ZONE AT IMPLANT).

Description of Event or Problem · 1

A TALENT STENT GRAFT WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN 8.7CM ABDOMINAL AORTIC ANEURYSM APPROX 32 MONTHS AGO. VESSEL MORPHOLOGY AT THE TIME OF IMPLANT WAS NOT REPORTED. AT THE TIME OF INITIAL IMPLANT, THE BIFURCATED STENT GRAFT IPSILATERAL LIMB WAS ONLY EXTENDED INTO THE ILIAC LIMB 1.5CM. IT WAS REPORTED THAT THE PT CAME IN FOR A ROUTINE F/U AND THE BIFURCATED STENT GRAFT IPSILATERAL LIMB HAD PULLED OUT OF THE ILIAC ARTERY AND RESULTED IN A DISTAL TYPE 1 ENDOLEAK. THE ANEURYSM WAS 8.2 CM TWO YEARS AGO AND CURRENTLY IS 9.2CM IN DIAMETER. THE PHYSICIAN ELECTED TO EXTEND THE STENT GRAFTS DOWN TO THE HYPOGASTRIC ARTERY WITH AN ILIAC LIMB AND THE ENDOLEAK WAS RESOLVED. NO ADD'L CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR NA V00101765

Patients

Seq Age Sex Outcome Treatment
1 80 YR Required Intervention