TALENT ABDOMINAL STENT GRAFT SYSTEM (CT)
Report
- Report Number
- 2953200-2011-00698
- Event Type
- Injury
- Date Received
- March 25, 2011
- Date of Event
- February 23, 2011
- Report Date
- February 23, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION, RESULTS: (STENT GRAFT WAS INITIALLY PLACED LOW INTENTIONALLY DUE TO THE ACCESSORY RENAL ARTERY), (ACCESSORY RENAL). CONCLUSIONS: (ACCESSORY RENAL), (STENT GRAFT WAS INITIALLY PLACED LOW INTENTIONALLY DUE TO THE ACCESSORY RENAL ARTERY).
A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROXIMATELY 33 MONTHS AGO. ANEURYSM AND VESSEL MORPHOLOGY FROM THE TIME OF IMPLANT WERE NOT REPORTED. IT WAS REPORTED THAT DURING A RECENT FOLLOW-UP APPROXIMATELY 1 MONTH AGO THE PROXIMAL PORTION OF THE BIFURCATED STENT GRAFT FABRIC WAS FOUND 15 MM BELOW THE RENAL ARTERIES AND THE SUPRA RENAL STENT APEXES WERE RIGHT AT THE LEVEL OF THE RENALS. THE PHYSICIAN DOES NOT BELIEVE THE STENT GRAFT MIGRATED AND THAT THIS IS WHERE THE STENT GRAFT WAS IMPLANTED. THERE WAS NO PROXIMAL TYPE 1 ENDOLEAK. THERE WAS HOWEVER, A DISTAL TYPE 1 ENDOLEAK THAT WAS COMING FROM THE LEFT (CONTRALATERAL) SIDE AND THIS WAS ATTRIBUTED TO A DILATED LEFT ILIAC ARTERY; THE DIAMETER OF THE ILIAC ARTERY IS UNKNOWN AND ITS LENGTH IS 2 CM LONG. IT WAS NOTED THAT THIS STENT GRAFT WAS NOT EXTENDED DOWN FAR ENOUGH TO THE LEVEL OF THE HYPOGASTRIC ARTERY AND CREATED A DISTAL TYPE 1 ENDOLEAK. THIS RESULTED IN ANEURYSM ENLARGEMENT (THE EXACT SIZE UNKNOWN), SEE MFR # 2953200-2011-00699. ON (B)(6) 2011, THE INTERNAL ILIAC ARTERY WAS COIL EMBOLIZED AND TWO DAYS LATER ENDURANT LIMBS WERE PLACED TO EXTEND DOWN TO THE HYPOGASTRIC ON THE CONTRALATERAL SIDE. AN ENDURANT CUFF WAS PLACED PROXIMALLY TO EXTEND TO THE RENAL ARTERIES. THERE IS A BRANCH OFF THE SMA THAT APPEARS TO BE AN ACCESSORY RENAL, SO THE GRAFT WAS PLACED INTENTIONALLY LOW AT THE TIME OF IMPLANT TO PRESERVE WHAT WAS THOUGHT TO BE AN ACCESSORY RENAL. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT ABDOMINAL STENT GRAFT SYSTEM (CT) | MIH | MEDTRONIC CARDIOVASCULAR | NA | V00114811 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Required Intervention |