FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 2032316 · Received March 23, 2011

Report

Report Number
2531779-2011-01861
Event Type
Injury
Date Received
March 23, 2011
Date of Event
February 11, 2011
Report Date
February 22, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Removal / Correction Number
2531779-02/25/11-001-R
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. THERE IS NO INVESTIGATION NECESSARY. CARTRIDGES WITH LOT #B201581 WERE CONFIRMED TO BE DEFECTIVE. WHEN TESTED, FLUID WAS OBSERVED LEAKING FROM THE PLUNGER END OF THE CARTRIDGE.

Description of Event or Problem · 1

THE REPORTER REPORTED THAT ON (B)(6), 2011, THE PT OBTAINED A BLOOD GLUCOSE READING OF 'GREATER THAN 200 MG/DL'. ON THAT DAY, THE PT NOTED THE INSULIN CARTRIDGE WAS LEAKING INTO THE PUMP'S CARTRIDGE COMPARTMENT. THE PT REPLACED THE CARTRIDGE AND HIS ELEVATED BLOOD GLUCOSE LEVELS RESOLVED. ON (B)(6), 2011, THE PT EXPERIENCED THE SYMPTOMS OF NAUSEA, VOMITING AND CHEST PAIN. THE PT WAS ADMITTED TO THE HOSPITAL FOR 24 HOURS OBSERVATION. TROUBLESHOOTING REVEALED THE PT'S TECHNIQUE WAS CORRECT, THERE WERE NO ISSUES WITH THE INFUSION SITE, THE TUBING AND CANNULA WERE CORRECT AND THE TOTAL DAILY DELIVERY AND PROGRAMMED WERE CORRECT AND MATCHED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP CARTRIDGE LZG ANIMAS CORPORATION IR 1200 / 1250 / 2020 / OTP B201581

Patients

Seq Age Sex Outcome Treatment
1 28 YR Hospitalization| L| R