ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2011-01868
- Event Type
- Injury
- Date Received
- March 23, 2011
- Date of Event
- February 11, 2011
- Report Date
- February 22, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Removal / Correction Number
- 2531779-02/25/11-001-R
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. THERE IS NO INVESTIGATION NECESSARY. CARTRIDGES WITH LOT #B201576 WERE CONFIRMED TO BE DEFECTIVE. WHEN TESTED, FLUID WAS OBSERVED LEAKING FROM THE PLUNGER END OF THE CARTRIDGE.
ON (B)(6) 2011, THE PATIENT'S FATHER CONTACTED ANIMAS ALLEGING THAT THE PATIENT'S BLOOD GLUCOSE (BG) HAS BEEN FLUCTUATING FOR THE PAST 3 MONTHS. THE REPORTER CLAIMED HE RECEIVED AN EMAIL FROM ANIMAS REGARDING A FIELD CORRECTION FOR LEAKING CARTRIDGES AND FEELS THAT THE PATIENT'S BG EXCURSION MAY HAVE BEEN AS A RESULT OF USING ONE OF THE AFFECTED LOTS. THE PATIENT'S FATHER STATED THAT ON (B)(6) 2011, THE PATIENT WAS TAKEN TO THE HOSPITAL FOR A HIGH BG. THE PATIENT'S FATHER CLAIMED THE PATIENT WAS TESTING IN 300-400 MG/DL, TESTED POSITIVE FOR KETONES AND HAD SYMPTOMS OF NAUSEA AND VOMITING. IT WAS REPORTED THAT THE PATIENT WAS KEPT ON THE PUMP WHILE AT THE HOSPITAL AND WAS TREATED WITH INJECTIONS AND DISCHARGED THE FOLLOWING DAY. AT THE TIME OF TROUBLESHOOTING, THE PUMP WAS NOT AVAILABLE FOR REVIEW. THE PATIENT'S FATHER REPORTED THAT HE HAS NOT NOTICED ANY WETNESS IN THE CARTRIDGE COMPARTMENT AND HAS NOT NOTED ANY LEAKS FROM THE CARTRIDGE. THE REPORTER CONFIRMED THE PATIENT CHANGES SITE EVERY OTHER DAY, DENIES ANY PROBLEMS WITH KINKED SITES, DENIED ANY ISSUE WITH INSERTION AND CONFIRMED THERE WERE NO AIR BUBBLES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP CARTRIDGE | LZG | ANIMAS CORP. | IR 1200/1250/2020/OTP | B201576 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR | Hospitalization| L| R |