DIMENSION(R) CLINICAL CHEMISTRY SYSTEM

MDR report key: 2032243 · Received March 29, 2011

Report

Report Number
2517506-2011-00054
Event Type
Malfunction
Date Received
March 29, 2011
Date of Event
March 15, 2011
Report Date
March 15, 2011
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Product Code
LEH
PMA / PMN Number
K061655
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED VANCOMYCIN RESULT IS SAMPLE INTEGRITY. THE IFU FOR THE DIMENSION VISTA VANCOMYCIN FLEX REAGENT CARTRIDGE CONTAINS THE FOLLOWING INFORMATION: "SPECIMENS SHOULD BE FREE OF PARTICULATE MATTER. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FALSELY DEPRESSED VANCOMYCIN RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAMPLE WAS REPEATED ON AN ALTERNATE DIMENSION VISTA(R) ANALYZER AND A HIGHER RESULT WAS OBTAINED AND REPORTED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED VANCOMYCIN RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIMENSION(R) CLINICAL CHEMISTRY SYSTEM VANCOMYCIN FLEX(R) REAGENT CARTRIDGE LEH SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW 10327BA