BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG
Report
- Report Number
- 1917413-2024-00907
- Event Type
- Malfunction
- Date Received
- September 27, 2024
- Date of Event
- August 30, 2024
- Report Date
- October 4, 2024
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- PJE
- UDI-DI
- 30382903627951
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
B3: DATE OF EVENT IS UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
INVESTIGATION SUMMARY: BD RECEIVED 13 SAMPLES FOR INVESTIGATION. THE SAMPLES ALONG WITH 100 RETENTION SAMPLES WERE VISUALLY INSPECTED: CUSTOMER RETURNED SAMPLES CONTAINED GEL AIR BUBBLES; NO FOREIGN MATTER(FM) WAS OBSERVED. 3 RETENTION SAMPLES CONTAINED GEL AIR BUBBLES; NO FM WAS OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE [GEL AIR BUBBLES. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR FM. BD WAS NOT ABLE TO IDENTIFY AN EXACT ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED PRIOR TO USE OF BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG AN UNSPECIFIED NUMBER OF TUBES HAD GEL AIR BUBBLES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
IT WAS REPORTED PRIOR TO USE OF BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG AN UNSPECIFIED NUMBER OF TUBES HAD GEL AIR BUBBLES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1685018 | BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG | BLOOD/PLASMA COLLECTION DEVICE FOR DNA TESTING | PJE | BECTON, DICKINSON & CO. (BROKEN BOW) | 4011397 | 30382903627951 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |