FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CXE1 PUMP UNREFURB

MDR report key: 2032192 · Received March 29, 2011

Report

Report Number
6000001-2011-02290
Event Type
Malfunction
Date Received
March 29, 2011
Date of Event
February 16, 2011
Report Date
February 16, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-12/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A 403:317:871:000 ERROR WAS CONFIRMED BUT NOT REPRODUCED. THE CAUSE OF THE REPORTED CONDITION WAS DETERMINED TO BE DEFECTIVE USER INTERFACE MODULE PRINTED CIRCUIT BOARD (UIM PCB). THE UIM PCB WAS REPLACED TO FIX THE REPORTED CONDITION. BAXTER HAS FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA (B)(4). A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

THE BAXTER SERVICE TECHNICIAN REPORTED A COLLEAGUE INFUSION PUMP WHICH EXPERIENCED FAILURE CODE 403:317:871:000 DURING DEVICE EVALUATION TESTING, WHICH INTERRUPTED DELIVERY. THIS DEVICE IS A REMEDIATED COLLEAGUE INFUSION PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.