BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG
Report
- Report Number
- 1917413-2024-00906
- Event Type
- Malfunction
- Date Received
- September 27, 2024
- Date of Event
- August 14, 2024
- Report Date
- August 30, 2024
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- PJE
- UDI-DI
- 30382903627951
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY: BD RECEIVED ONE PHOTO FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR FOREIGN MATTER(FM) WAS OBSERVED. THERE IS A WHITE PARTICLE ON THE INSIDE OF THE TUBE. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE FM. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED PRIOR TO USE OF BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG, 1 TUBE CONTAINED FOREIGN MATTER(FM) - 2 WHITE LUMPS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1371772 | BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG | BLOOD/PLASMA COLLECTION DEVICE FOR DNA TESTING | PJE | BECTON, DICKINSON & CO. (BROKEN BOW) | 4073675 | 30382903627951 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |