COLLEAGUE VOLUMETRIC INFUSOR PUMP-SINGLE CHANNEL
Report
- Report Number
- 6000001-2011-02274
- Event Type
- Malfunction
- Date Received
- March 28, 2011
- Date of Event
- January 19, 2011
- Report Date
- January 20, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Removal / Correction Number
- 6000001-12/1/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE REPORTED CONDITION WAS CONFIRMED BUT NOT DUPLICATED. THE ASSIGNABLE CAUSE WAS DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND BATTERY HARNESS HAVE BEEN REPLACED. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). A SERVICE HISTORY REVIEW HAS BEEN PERFORMED AND THE DEVICE HAS BEEN PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. A TREND REVIEW HAS BEEN PERFORMED AND THERE HAVE BEEN SIMILAR REPORTS MADE TO BAXTER. THE ROOT CAUSE WILL CONTINUE TO BE INVESTIGATED THROUGH CAPA (CORRECTIVE AND PREVENTIVE ACTION) MDQ-CAPA (B)(4).
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A MALFUNCTION OF DAMAGED BATTERY. THE EVENT WAS REPORTED TO HAVE OCCURRED DURING PROGRAMMING/SETUP. ACCORDING TO THE HOSPITAL REPRESENTATIVE THE PATIENT WAS NOT INVOLVED. NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. DURING THE PRODUCT EVALUATION AT BAXTER, IT WAS DISCOVERED THAT A BATTERY DEPLETED SET ALARM OCCURRED DURING INFUSION, WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 5.09.90, CATEGORIZED AS REMEDIATED.