GN RESOUND
Report
- Report Number
- 3005650109-2024-00077
- Event Type
- Malfunction
- Date Received
- September 27, 2024
- Date of Event
- August 26, 2024
- Report Date
- September 26, 2024
- Manufacturer
- GN HEARING A/S
- Product Code
- OSM
- UDI-DI
- 05708296260638
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
MANUFACTURER'S REF# (B)(4). MANUFACTURER'S INVESTIGATION: TECHNICAL INVESTIGATION CONCLUDED: A DHR REVIEW HAVE BEEN COMPLETED. NO DEVIATION OR CHANGES WERE FOUND DURING MANUFACTURING OF THE DEVICE. CLINICAL CONCLUSION: IT HAS BEEN REPORTED THAT THE WAX GUARD GROMMET/BRASS RING HAS FALLEN OUT OF AN IN THE EAR HEARING AID. THE GROMMET DID NOT FALL OUT IN THE USER'S EAR CANAL. SINCE THE GROMMET FELL OUT, USER IS UNABLE TO USE A WAX GUARD ON THE DEVICE. THERE WAS NO HARM TO THE USER FOLLOWING THE INCIDENT. CLINICAL ASSESSMENT IS THAT THE EVENT DID NOT CAUSE HARM TO THE USER. HEARING AIDS NEED TO HAVE DETACHABLE SPARE PARTS AS WAX FILTERS. THEREFORE, ELIMINATING THE RISK OF A SPARE PARTS REMAINING IN THE EAR CANAL IS NOT POSSIBLE. A WAX FILTER CAN BECOME DETACHED AND REMAIN IN THE USER'S EAR CANAL WHEN THE HEARING AID IS REMOVED OR IF THE CERUSTOP BUSHING HAS BECOME DISLODGED FROM EXCESSIVE FORCE APPLIED DURING REPLACEMENTS OR CLEANING OF THE WAX FILTER. WITHOUT BUSHING, A REPLACED CERUSTOP IS NOT SECURE AND MAY FALL OUT IN THE EAR. THE WAX FILTER THEN CONSTITUTES A FOREIGN BODY IN THE EAR CANAL AND SHOULD BE REMOVED AS SOON AS POSSIBLE AS IT CAN POSE AN INFECTION RISK. THE HAZARD IS IDENTIFIED IN THE RISK ANALYSIS AND MITIGATED ON DEVICE DESIGN BY ENSURING THAT THE WAX FILTER CANNOT BE REMOVED WITHOUT A TOOL. THE USER GUIDE INCLUDES AN INSTRUCTION ON HOW TO REPLACE A WAX FILTER WITH A TOOL AND TO CONTACT THE HCP IF A FOREIGN OBJECT IS LOCATED IN THE EAR CANAL. RISK ASSESSMENT: RISKS RELATED TO MECHANICAL DAMAGE / LOOSE PARTS ARE GENERALLY KNOWN, CONSIDERED IN THE RISK ANALYSIS, MITIGATED AND COMMUNICATED TO THE USER. THIS RISK IS DEEMED TO BE OF AN ACCEPTABLE NATURE. DEVICE INVESTIGATION CONCLUDED: NO DEVICE INVESTIGATION PERFORMED AS THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER. MANUFACTURER'S INVESTIGATION IS FINAL. THIS IS A COMBINED (INITIAL AND FINAL) REPORT.
IT WAS REPORTED THAT THE WAX GUARD GROMMET HAS FALLEN OUT OF AN IN THE EAR HEARING AID. THE GROMMET DID NOT FALL OUT INSIDE THE USER'S EAR CANAL. NOTHING WAS STUCK IN THE USER'S EAR, USER DID NOT SUSTAIN ANY HARM FROM THE EVENT. NO FURTHER FOLLOW UP IS AVAILABLE OR EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1684899 | GN RESOUND | HEARING AID, , AIR CONDUCTION WITH WIRELESS TECHNOLOGY | OSM | GN HEARING A/S | RU9ITE-DWC-MP | 05708296260638 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Male |