FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX WINKELSTUCK 25LPA
MDR report key: 2031992
·
Received March 17, 2011
Report
- Report Number
- 3003637274-2011-00015
- Event Type
- Injury
- Date Received
- March 17, 2011
- Date of Event
- January 1, 2011
- Report Date
- February 18, 2011
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
DURING A STANDARD PROCEDURE, A DENTAL ELECTRICAL HANDPIECE GOT HOT AND BURNED THE PT. THE LOCATION OF THE BURN IS NOT KNOW. PT RECEIVED SOME SAMPLES OF OINTMENT AND MEDICATION. THE ANALYSIS DID SHOW THAT THE HEAD IS DENTED AT THE BACKCAP. HENCE THE BACK CAP BUTTON STICKS IN AND CAUSES THE HEAD TO HEAT UP. EXEMPTION NUMBER (B)(4). KAVO DENTAL (B)(4) (THE MFR) IS SUBMITTING THE REPORT ON BEHALF OF KAVO DENTAL (B)(4) (THE IMPORTER).
Description of Event or Problem · 1
DURING A STANDARD PROCEDURE, A DENTAL ELECTRICAL HANDPIECE GOT HOT AND BURNED THE PT. THE LOCATION OF THE BURN IS NOT KNOW. PT RECEIVED SOME SAMPLES OF OINTMENT AND MEDICATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX WINKELSTUCK 25LPA | DENTAL HANDPIECE | EFB | KAVO DENTAL GMBH | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |