FDA Adverse Event Injury Summary report: N

GENTLEPOWER LUX WINKELSTUCK 25LPA

MDR report key: 2031992 · Received March 17, 2011

Report

Report Number
3003637274-2011-00015
Event Type
Injury
Date Received
March 17, 2011
Date of Event
January 1, 2011
Report Date
February 18, 2011
Manufacturer
KAVO DENTAL GMBH
Product Code
EFB
PMA / PMN Number
K073478
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

DURING A STANDARD PROCEDURE, A DENTAL ELECTRICAL HANDPIECE GOT HOT AND BURNED THE PT. THE LOCATION OF THE BURN IS NOT KNOW. PT RECEIVED SOME SAMPLES OF OINTMENT AND MEDICATION. THE ANALYSIS DID SHOW THAT THE HEAD IS DENTED AT THE BACKCAP. HENCE THE BACK CAP BUTTON STICKS IN AND CAUSES THE HEAD TO HEAT UP. EXEMPTION NUMBER (B)(4). KAVO DENTAL (B)(4) (THE MFR) IS SUBMITTING THE REPORT ON BEHALF OF KAVO DENTAL (B)(4) (THE IMPORTER).

Description of Event or Problem · 1

DURING A STANDARD PROCEDURE, A DENTAL ELECTRICAL HANDPIECE GOT HOT AND BURNED THE PT. THE LOCATION OF THE BURN IS NOT KNOW. PT RECEIVED SOME SAMPLES OF OINTMENT AND MEDICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTLEPOWER LUX WINKELSTUCK 25LPA DENTAL HANDPIECE EFB KAVO DENTAL GMBH 25LPA

Patients

Seq Age Sex Outcome Treatment
1 Other