FDA Adverse Event Malfunction Summary report: N

BASAL/BOLUS INFUSOR 2ML X 2ML6 PACK

MDR report key: 2031985 · Received March 28, 2011

Report

Report Number
6000001-2011-02257
Event Type
Malfunction
Date Received
March 28, 2011
Date of Event
March 9, 2011
Report Date
March 10, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE CAUSE IS A LOOSE WINGED LUER CAP. ADDITIONAL: A BATCH REVIEW HAS BEEN PERFORMED AND THERE WERE NO EXCEPTIONS NOTED DURING THE MANUFACTURING OF THIS LOT.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER (B)(4) TO REPORT AN INFUSOR IN WHICH THERE WAS LEAKAGE OBSERVED IN THE DISPENSER BAG. THE EVENT OCCURRED DURING STORAGE. THE FACILITY ATTEMPTED TO FASTEN THE BLUE WINGED CAP AGAIN AND STORED THE DEVICE, BUT LEAKAGE WAS OBSERVED AGAIN. THE DEVICE WAS FILLED WITH MORPHINE HYDROCHLORIDE AND SALINE. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BASAL/BOLUS INFUSOR 2ML X 2ML6 PACK PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 10H015

Patients

Seq Age Sex Outcome Treatment
1