VOYAGER RX CORONARY DILATATION CATHETER
Report
- Report Number
- 2024168-2011-02077
- Event Type
- Malfunction
- Date Received
- March 28, 2011
- Date of Event
- March 2, 2011
- Report Date
- March 3, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- LOX
- PMA / PMN Number
- P810046
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE CUSTOMER REPORTED THAT THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS PROVIDED. REVIEW OF THE DEVICE HISTORY RECORD IS FORTHCOMING. A FOLLOW-UP WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.
(B)(4). BALLOON MATERIAL RUPTURES CAN BE AFFECTED BY NUMEROUS FACTORS INCLUDING, BUT ARE NOT LIMITED TO, BALLOON DAMAGE DURING MANUFACTURING, MATERIALS, HANDLING, LESION CALCIFICATION AND TORTUOSITY, AN INTERACTION WITH A PREVIOUSLY IMPLANTED STENT, INSUFFICIENT PREPARATION PRIOR TO USE OR FROM INTERACTIONS WITH OTHER DEVICES. SINCE THERE WAS NO REPORT OF ANY LEAKS NOTED DURING PREPARATION FOR USE, THIS MAY INDICATE THAT THE BALLOON WAS NOT DAMAGED PRIOR TO THE PROCEDURE. ADDITIONALLY, THE LESION WAS MILDLY TORTUOUS, MILDLY CALCIFIED AND 90% STENOSED, WHICH MAY HAVE CONTRIBUTED TO THE REPORTED DIFFICULTY. IT IS POSSIBLE THAT THE BALLOON MATERIAL WAS DAMAGED (SCRATCHED) FROM INTERACTIONS WITH OTHER DEVICES AND/OR THE CALCIFIED LESION SUCH THAT UPON INFLATION ATTEMPT, THE BALLOON RUPTURED. ULTIMATELY, RETURN OF THE VOYAGER MAY HAVE AIDED THE INVESTIGATION IN DETERMINING A CAUSE FOR THE EXPERIENCED RUPTURE. A REVIEW OF THE FINISHED PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY NONCONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT AND ALL LOT RELEASE TESTING MET MANUFACTURING CRITERIA. FURTHERMORE, A QUERY OF THE COMPLAINT-HANDLING DATABASE DID NOT REVEAL ANY OTHER INCIDENTS REPORTED FOR A BALLOON RUPTURE FROM THIS LOT, SUGGESTING THAT THERE ARE NO LOT SPECIFIC PRODUCT QUALITY DEFICIENCIES. BASED ON THE INFORMATION RECEIVED WITH THIS COMPLAINT AND WITHOUT THE PRODUCT TO EXAMINE, A DEFINITIVE CAUSE FOR THE REPORTED BALLOON RUPTURE CANNOT BE DETERMINED. ALL DILATATION CATHETERS ARE VISUALLY INSPECTED AND LEAK TESTED DURING THE MANUFACTURING PROCESS. A SAMPLING OF UNITS IS ALSO DESTRUCTIVELY TESTED TO VERIFY RBP AND BALLOON INTEGRITY.
IT WAS REPORTED THAT THE TARGET LESION WAS THE DISTAL RIGHT CORONARY ARTERY (RCA) WITH MILD TORTUOSITY, MILD CALCIFICATION AND 90% STENOSIS. A GUIDE WIRE CROSSED THE LESION AND THE VOYAGER RX BALLOON CATHETER WAS ADVANCED TO THE LESION; HOWEVER, WHEN THE BALLOON WAS INFLATED FOR THE FIRST TIME IT RUPTURED AT 7 ATMOSPHERE IMMEDIATELY AFTER PRESSURIZATION. A 2.5 X 15 MM NON-ABBOTT BALLOON CATHETER WAS USED TO FURTHER DILATE THE LESION. THERE WAS NO ADVERSE PATIENT EFFECTS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VOYAGER RX CORONARY DILATATION CATHETER | CORONARY DILATATION CATHETER | LOX | AV-TEMECULA-CT | 0102562 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GUIDE WIRE: RUNTHROUGH, GUIDE CATH: HEARTRAIL |