FDA Adverse Event Malfunction Summary report: N

HOMECHOICE LOW RECIRCULATION VOLUME APD SET WITH CASSETTE

MDR report key: 2031946 · Received March 28, 2011

Report

Report Number
1423500-2011-03753
Event Type
Malfunction
Date Received
March 28, 2011
Date of Event
March 5, 2011
Report Date
March 5, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K012988
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS DISCARDED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON THE COMPLETION OF BAXTER'S QUALITY REVIEW.

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT WAS NOT CONFIRMED. BASED ON THE INFORMATION GATHERED DURING BAXTER?S INVESTIGATION, THE CAUSE OF THE SYSTEM ERROR 2240 WAS NOT DETERMINED. THE LOT NUMBER IS UNKNOWN; THEREFORE, A BATCH REVIEW WAS NOT PERFORMED. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER TO REPORT A SYSTEM ERROR (SE) 2240 AND 2367 WHICH OCCURRED ON HOME CHOICE (HC) DURING USE DURING DRAIN 7 OF 7. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM AND HAD THE CARE GIVER (CG) CLOSE CLAMPS AND THEN CYCLE POWER TO CLEAR BOTH SE 2240 AND SE 2367. THE TSR ADVISED THE CG TO FINISH THE THERAPY WITH MANUALS. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE LOW RECIRCULATION VOLUME APD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE