PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2011-02069
- Event Type
- Injury
- Date Received
- March 28, 2011
- Date of Event
- February 25, 2011
- Report Date
- March 3, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: EVALUATION OF THE RETURNED DEVICE FOUND THAT IT WAS PARTIALLY DEPLOYED. BOTH CUFFS HAD BEEN CAPTURED AND THE POSTERIOR CUFF WAS ATTACHED TO THE NEEDLE TIP. HOWEVER, THE LINK WAS PULLED FROM THE POSTERIOR CUFF DURING SUTURE HARVESTING. A LINK BREAK DURING NEEDLE RETRIEVAL AND A CUFF MISS HAVE THE SAME RESULT AND APPEARANCE DURING DEPLOYMENT; THEREFORE, THE REPORTED EXPERIENCE IS CONFIRMED. THE OTHER COMPONENTS WERE NORMAL AND UNDAMAGED. AFTER NEEDLE DEPLOYMENT, THE SUTURE IS HARVESTED AND PULLED THROUGH THE SUTURE BEARING AND ANTERIOR NEEDLE GUIDE; RESISTANCE AT THIS POINT OF DEPLOYMENT CAN CAUSE THE LINK TO DETACH. A LINK DETACHMENT CAN BE INFLUENCED BY MANY FACTORS INCLUDING, BUT NOT LIMITED TO, MANUFACTURING, FAILURE TO POSITION AND MAINTAIN THE DEVICE AT A 45 DEGREE ANGLE THROUGHOUT DEPLOYMENT, ROTATING THE DEVICE AT ANY POINT DURING PLUNGER/NEEDLE DEPLOYMENT, DEPLOYMENT IN CHALLENGING ANATOMIES (E.G. HEAVY CALCIFIED ARTERIES, MORBIDLY OBESE PATIENTS, ETC.), AGGRESSIVELY DEPLOYING OR REMOVING THE PLUNGER AND/OR INADEQUATE POSITIONING OF THE FOOT AGAINST THE ARTERIAL WALL. BASED ON THE ANALYSIS THE PROBABLE CAUSE OF THE LINK DETACHMENT IS RELATED TO THE OPERATIONAL CONTEXT DURING USE. THERE WAS NO PRODUCT QUALITY DEFICIENCY DETECTED THAT WOULD CONTRIBUTE TO THIS EVENT. A REVIEW OF THE FINISHED DEVICE LOT HISTORY RECORDS DID NOT REVEAL ANY NONCONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT. A QUERY OF THE COMPLAINT-HANDLING DATABASE WAS PERFORMED. THERE DOES NOT APPEAR TO BE ANY INDICATION OF A LOT SPECIFIC PRODUCT QUALITY DEFICIENCY. EACH COMPONENT IS INSPECTED DURING MANUFACTURING AND EACH FINISHED GOODS LOT IS INSPECTED BY SAMPLING.
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. ESTIMATE AGE. PATIENT WAS REPORTED TO BE (B)(6). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
IT WAS REPORTED THAT A TECHNICIAN TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF A RIGHT COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, UPON PLUNGER REMOVAL, THE SUTURE KEPT PULLING OUT. A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THOUGH REQUESTED, ADDITIONAL INFORMATION WAS NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 950386H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Required Intervention |