PROXIMATE ILS CURVED CIR STAPL
Report
- Report Number
- 3005075853-2011-01220
- Event Type
- Malfunction
- Date Received
- March 28, 2011
- Date of Event
- March 2, 2011
- Report Date
- March 7, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K983536
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). DAMAGED ANCILLARY TROCAR. THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION AND WITH THE TIP OF THE ANCILLARY TROCAR LODGED INSIDE THE ANVIL. THE BREAKAWAY WASHER WAS PRESENT AND UNCUT AND THE DEVICE WAS RECEIVED FULLY LOADED WITH STAPLES. THE DEVICE WAS TESTED FOR FUNCTIONALITY USING A TEST ANVIL AND WAS FIRED AT THE HIGH-B SETTING; IT FIRED AND FORMED ALL THE STAPLES, AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. IT COULD NOT BE ASCERTAINED AS TO HOW THE DAMAGE TO THE ANCILLARY TROCAR OCCURRED. IT SHOULD BE NOTED THAT IF TORQUE OR EXCESS FORCE IS APPLIED TO THE ANCILLARY TROCAR, IT MAY CAUSE THE TROCAR TO BREAK. IT SHOULD BE NOTED THAT TO DETACH THE ANCILLARY TROCAR, IT SHOULD BE ROTATED 45 DEGREES IN THE ANVIL SHAFT SO THAT THE FINGER NOTCHES ARE PERPENDICULAR TO THE LOCKING SPRINGS (UNLOCKED POSITION). UTILIZING THE FINGER NOTCHES, PULL THE ANCILLARY TROCAR OUT OF THE ANVIL SHAFT. PLEASE REFERENCE THE INSTRUCTIONS FOR USE FOR ADDITIONAL INFORMATION. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING A TOTAL GASTRECTOMY PROCEDURE, THE DEVICE WAS USED FOR ROUX-EN-Y ANASTOMOSIS AT THE JEJUNUM. THE ANCILLARY BLUE TROCAR WAS NOT PULLED OUT FROM THE ANVIL SHAFT. THE DOCTOR DELUDED HIMSELF THAT THE ANCILLARY TROCAR WOULD COME OFF WHEN THE LOCKING SPRING WAS PUSHED. THEREFORE, HE HELD THE LOCKING SPRING TIGHTLY WITH A FORCEPS SUCH AS "PEAN." FINALLY, THE ANCILLARY TROCAR WAS BROKEN AND GOT STUCK IN THE ANVIL. THE TARGET TISSUE WAS SUTURED BY HAND AND THE OPERATION WAS FINISHED WITHOUT ANY PROBLEM. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. ADDITIONAL INFORMATION: THE EVENT OCCURRED BEFORE USE ON THE PATIENT, SO THE DEVICE WAS NOT USED FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROXIMATE ILS CURVED CIR STAPL | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | G4V881 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |