FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2031829 · Received March 21, 2011

Report

Report Number
2531779-2011-01787
Event Type
Malfunction
Date Received
March 21, 2011
Report Date
February 21, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. DURING INVESTIGATION, IT WAS FOUND THAT ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTION. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, DURING EVALUATION, IT WAS FOUND THAT THE DISPLAY HAD A REDDISH TINT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD IS INTERMITTENTLY RESPONDING. IT WAS REPORTED THAT THERE IS WATER EXPOSURE TO THE PUMP IN THE SHOWER AND THERE IS NO DAMAGE TO KEYPAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1