FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2031829
·
Received March 21, 2011
Report
- Report Number
- 2531779-2011-01787
- Event Type
- Malfunction
- Date Received
- March 21, 2011
- Report Date
- February 21, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. DURING INVESTIGATION, IT WAS FOUND THAT ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTION. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, DURING EVALUATION, IT WAS FOUND THAT THE DISPLAY HAD A REDDISH TINT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD IS INTERMITTENTLY RESPONDING. IT WAS REPORTED THAT THERE IS WATER EXPOSURE TO THE PUMP IN THE SHOWER AND THERE IS NO DAMAGE TO KEYPAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |