FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2031823 · Received March 21, 2011

Report

Report Number
2531779-2011-01780
Event Type
Malfunction
Date Received
March 21, 2011
Report Date
February 21, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE DOWN KEYPAD BUTTON WAS INTERMITTENTLY RESPONSIVE DURING TESTING. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER THE DOWN ARROW AND OK BUTTONS. UNRELATED TO THE COMPLAINT, DURING EVALUATION A CRACKED BATTERY COMPARTMENT WAS OBSERVED, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. THE USER GUIDE WARNS THAT CRACKS,CHIPS, OR DAMAGE TO THE PUMP MAY IMPACT THE BATTERY CONTACT AND / OR THE WATERPROOF FEATURE OF THE PUMP. IT WAS ALSO OBSERVED THAT THE DISPLAY SCREEN HAD A DIM REDDISH TINT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DOWN ARROW BUTTON IS SLOW TO RESPOND. IT WAS REPORTED THE KEYPAD IS INTACT AND THE PUMP IS NOT EXPOSED TO MOISTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1