FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2031818
·
Received March 21, 2011
Report
- Report Number
- 2531779-2011-01785
- Event Type
- Malfunction
- Date Received
- March 21, 2011
- Report Date
- February 21, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE DOWN KEYPAD BUTTON WAS FOUND TO BE HYPERSENSITIVE WHEN PRESSED. DURING INVESTIGATION, THE DOWN ARROW KEY CONTACT WAS OBSERVED TO BE MISALIGNED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER THE DOWN ARROW BUTTON.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DOWN ARROW BUTTON IS VERY SENSITIVE AND IS INTERMITTENTLY RESPONDING. IT WAS ALSO REPORTED THERE ARE NO TEARS OR PEELING OF THE KEYPAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |