FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2031818 · Received March 21, 2011

Report

Report Number
2531779-2011-01785
Event Type
Malfunction
Date Received
March 21, 2011
Report Date
February 21, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE DOWN KEYPAD BUTTON WAS FOUND TO BE HYPERSENSITIVE WHEN PRESSED. DURING INVESTIGATION, THE DOWN ARROW KEY CONTACT WAS OBSERVED TO BE MISALIGNED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER THE DOWN ARROW BUTTON.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DOWN ARROW BUTTON IS VERY SENSITIVE AND IS INTERMITTENTLY RESPONDING. IT WAS ALSO REPORTED THERE ARE NO TEARS OR PEELING OF THE KEYPAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1