FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2031790
·
Received March 21, 2011
Report
- Report Number
- 2531779-2011-01740
- Event Type
- Malfunction
- Date Received
- March 21, 2011
- Report Date
- February 21, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. DURING INVESTIGATION, IT WAS FOUND THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, DURING EVALUATION, IT WAS ALSO OBSERVED THAT THE DISPLAY HAD A REDDISH TINT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD IS NOT RESPONDING APPROPRIATELY. IT WAS REPORTED THAT THE PUMP IS NOT EXPOSED TO MOISTURE AND THE KEYPAD AND PUMP ARE INTACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |