FDA Adverse Event Malfunction Summary report: N

AFFINITY 3

MDR report key: 2031767 · Received March 21, 2011

Report

Report Number
1824206-2011-01716
Event Type
Malfunction
Date Received
March 21, 2011
Date of Event
February 24, 2011
Report Date
February 24, 2011
Manufacturer
HILL-ROM INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACCOUNTS MAINTENANCE BELIEVES THE CASTERS ARE WORN FROM AGE. REPAIRS HAVE NOT BEEN COMPLETED.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED WHEN THEY ENGAGE THE BRAKE PEDAL, THE BRAKE DOES NOT HOLD IF THEY LEAN ON THE BED FRAME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 3 BIRTHING BED HDD HILL-ROM INC. 3700

Patients

Seq Age Sex Outcome Treatment
1