FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 2031742
·
Received March 21, 2011
Report
- Report Number
- 1824206-2011-01710
- Event Type
- Malfunction
- Date Received
- March 21, 2011
- Date of Event
- March 2, 2011
- Report Date
- March 2, 2011
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN FOUND THAT THE INTELIDRIVE WOULD NOT RUN DUE TO BAD PCM BOARD. CORROSION ON BOARD. REPLACED PCM BOARD TO RESOLVE THIS ISSUE. BED WAS FOUND IN BASEMENT SHOP.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT THE INTELIDRIVE WOULD NOT RUN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | P1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |