FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2031742 · Received March 21, 2011

Report

Report Number
1824206-2011-01710
Event Type
Malfunction
Date Received
March 21, 2011
Date of Event
March 2, 2011
Report Date
March 2, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND THAT THE INTELIDRIVE WOULD NOT RUN DUE TO BAD PCM BOARD. CORROSION ON BOARD. REPLACED PCM BOARD TO RESOLVE THIS ISSUE. BED WAS FOUND IN BASEMENT SHOP.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT THE INTELIDRIVE WOULD NOT RUN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. P1900

Patients

Seq Age Sex Outcome Treatment
1