FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 2031510 · Received March 21, 2011

Report

Report Number
9615739-2011-00073
Event Type
Malfunction
Date Received
March 21, 2011
Date of Event
February 24, 2011
Report Date
February 24, 2011
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE LEFT SIDERAIL INTERCABLE WAS PINCHED OFF AT THE SIDECOM CABLE FOR THE CALL SYSTEM. HE REPLACED THE INTERCABLE TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION REC'D INDICATES NO NURSE CALL IS ABLE TO BE PLACED FROM THE BED. THE FACILITY DOES HAVE A BACKUP HAND HELD THAT WAS WORKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAREASSIST BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1 UNK