FDA Adverse Event
Malfunction
Summary report: N
M SERIES
MDR report key: 2031457
·
Received March 18, 2011
Report
- Report Number
- 1220908-2011-00672
- Event Type
- Malfunction
- Date Received
- March 18, 2011
- Report Date
- February 25, 2011
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- PMA / PMN Number
- K990762
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
ZOLL MEDICAL CORPORATION EVALUATED THE DEVICE AND THE REPORTED MALFUNCTION WAS NOT REPLICATED OR CONFIRMED. THE DEVICE WAS PUT THROUGH EXTENSIVE TESTING, WHICH INCLUDED ENVIRONMENTAL AND FUNCTIONAL TESTING, WITHOUT DUPLICATING THE MALFUNCTION. TESTING OF THE DEVICE WAS NOT ABLE TO DUPLICATE THE PROBLEM CONDITION. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. NO TREND IS ASSOCIATED WITH REPORTS OF THIS TYPE.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PT, THE DEVICE WAS UNABLE TO PACE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PT DUE TO THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | M SERIES | DEFIBRILLATOR/PACEMAKER | MKJ | ZOLL MEDICAL CORPORATION | MSERIES BI-PHASIC | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |