ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2031455
·
Received March 17, 2011
Report
- Report Number
- 2531779-2011-01763
- Event Type
- Malfunction
- Date Received
- March 17, 2011
- Date of Event
- February 17, 2011
- Report Date
- February 17, 2011
- Manufacturer
- ANIMAS CORP
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS. EVAL REVEALED CONTAMINATION UNDER BUTTON CONTACTS. THE OK BUTTON CONTACT WAS MISALIGNED. THE OK BUTTON DID NOT ACTIVATE DESIRED PUMP FUNCTIONS WHEN PRESSED. THE DOWN ARROW BUTTON INTERMITTENTLY ACTIVATED DESIRED PUMP FUNCTIONS. THE OK BUTTON DID NOT SPRING BACK WHEN PRESSED.
Description of Event or Problem · 1
THE PT STATED THAT THE OK BUTTON DOES NOT ACTIVATE DESIRED PUMP FUNCTIONS WHEN PRESSED. THE PUMP WAS NOT EXPOSED TO ANY WATER OR CLEANING PRODUCTS. SHE SAID THE OK BUTTON ALSO FELT FLAT. NO OTHER BUTTONS REPORTEDLY RESPOND TO BUTTON PRESSES.