FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2031455 · Received March 17, 2011

Report

Report Number
2531779-2011-01763
Event Type
Malfunction
Date Received
March 17, 2011
Date of Event
February 17, 2011
Report Date
February 17, 2011
Manufacturer
ANIMAS CORP
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. EVAL REVEALED CONTAMINATION UNDER BUTTON CONTACTS. THE OK BUTTON CONTACT WAS MISALIGNED. THE OK BUTTON DID NOT ACTIVATE DESIRED PUMP FUNCTIONS WHEN PRESSED. THE DOWN ARROW BUTTON INTERMITTENTLY ACTIVATED DESIRED PUMP FUNCTIONS. THE OK BUTTON DID NOT SPRING BACK WHEN PRESSED.

Description of Event or Problem · 1

THE PT STATED THAT THE OK BUTTON DOES NOT ACTIVATE DESIRED PUMP FUNCTIONS WHEN PRESSED. THE PUMP WAS NOT EXPOSED TO ANY WATER OR CLEANING PRODUCTS. SHE SAID THE OK BUTTON ALSO FELT FLAT. NO OTHER BUTTONS REPORTEDLY RESPOND TO BUTTON PRESSES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 60 YR