FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM

MDR report key: 2031442 · Received March 28, 2011

Report

Report Number
3005075853-2011-01211
Event Type
Malfunction
Date Received
March 28, 2011
Date of Event
March 4, 2011
Report Date
March 4, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE VISUAL INDICATOR WAS OVER-TRAVELED. WHEN THIS OCCURS, THE INDICATOR WHEEL MAY CONTINUE TO ROTATE AFTER INDICATION OF A COMPLETELY FIRED DEVICE. PLEASE NOTE THE OVER-TRAVEL OF THE INDICATOR WHEEL IS NOT RELATED TO THE REPORTED EVENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE FIRING WAS NOT SMOOTH AND THE DEVICE WAS NOT DEPLOYING GOOD CLIPS. THE CLIPS WERE NOT CLOSING OR SCISSORING ON EVERY FIRE. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO IMPACT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA H4321U

Patients

Seq Age Sex Outcome Treatment
1