ROTATING MULTIPLE CLIP APPLIER
Report
- Report Number
- 3005075853-2011-01209
- Event Type
- Malfunction
- Date Received
- March 28, 2011
- Date of Event
- March 3, 2011
- Report Date
- March 3, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT (A) WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE EJECTED THE REMAINING CLIPS. FINALLY, IT LOCKED OUT AS INTENDED. HOWEVER, IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE FOUND EJECTED CLIPS. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. THE ANALYSIS RESULTS FOUND THAT THE DEVICE (B) WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE EJECTED THE REMAINING CLIPS; NO MALFORMED CLIPS WERE NOTED. HOWEVER, THE LOCKOUT MECHANISM WAS FOUND TO BE NON-FUNCTIONAL. IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE EJECTED CLIPS. IN ORDER TO EVALUATE THE CONDITION OF THE INTERNAL COMPONENTS OF THE DEVICE, IT WAS DISASSEMBLED. UPON DISASSEMBLING, THE LOCKOUT POSTS WERE FOUND TO BE BROKEN WHICH SUGGEST THAT A LOCKOUT PREMATURE OCCURRED. PLEASE NOTE THAT THESE CONDITIONS ARE UNRELATED WITH THE INCIDENT REPORTED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. DEVICE B ADDITIONAL INFORMATION: BATCH # F9GY96 EXPIRATION DATE: 06/2014 MANUFACTURING DATE: 07/2009 (B)(4).
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIP WAS CONTINUOUSLY MISFIRED AND THE CLIP WAS BENT. UNKNOWN HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ROTATING MULTIPLE CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |