FDA Adverse Event Malfunction Summary report: N

ROTATING MULTIPLE CLIP APPLIER

MDR report key: 2031440 · Received March 28, 2011

Report

Report Number
3005075853-2011-01209
Event Type
Malfunction
Date Received
March 28, 2011
Date of Event
March 3, 2011
Report Date
March 3, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT (A) WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE EJECTED THE REMAINING CLIPS. FINALLY, IT LOCKED OUT AS INTENDED. HOWEVER, IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE FOUND EJECTED CLIPS. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. THE ANALYSIS RESULTS FOUND THAT THE DEVICE (B) WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE EJECTED THE REMAINING CLIPS; NO MALFORMED CLIPS WERE NOTED. HOWEVER, THE LOCKOUT MECHANISM WAS FOUND TO BE NON-FUNCTIONAL. IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE EJECTED CLIPS. IN ORDER TO EVALUATE THE CONDITION OF THE INTERNAL COMPONENTS OF THE DEVICE, IT WAS DISASSEMBLED. UPON DISASSEMBLING, THE LOCKOUT POSTS WERE FOUND TO BE BROKEN WHICH SUGGEST THAT A LOCKOUT PREMATURE OCCURRED. PLEASE NOTE THAT THESE CONDITIONS ARE UNRELATED WITH THE INCIDENT REPORTED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. DEVICE B ADDITIONAL INFORMATION: BATCH # F9GY96 EXPIRATION DATE: 06/2014 MANUFACTURING DATE: 07/2009 (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIP WAS CONTINUOUSLY MISFIRED AND THE CLIP WAS BENT. UNKNOWN HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ROTATING MULTIPLE CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1