FDA Adverse Event Injury Summary report: N

ANEURX STENT GRAFT WITH XCELERANT DELIVERY SYSTEM

MDR report key: 2031343 · Received March 23, 2011

Report

Report Number
2953200-2011-00684
Event Type
Injury
Date Received
March 23, 2011
Date of Event
February 22, 2011
Report Date
February 22, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL, RESULTS: (MIGRATION, ENDOLEAK), (DISEASE PROGRESSION WITH AORTIC NECK DILATATION). CONCLUSION: (DISEASE PROGRESSION WITH AORTIC NECK DILATATION).

Description of Event or Problem · 1

AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROX 64 MONTHS AGO. VESSEL MORPHOLOGY FROM THE TIME OF IMPLANT IS UNK. CURRENTLY THE ANEURYSM HAS INCREASED IN SIZE FROM 5.4 CM AT IMPLANT TO 5.8 CM AT THE TIME OF MIGRATION. THE AORTIC NECK WAS ANGULATED 15 DEGREES AND HAD DILATED FROM 23 MM IN DIAMETER AT THE TIME OF IMPLANT TO 25 MM, DUE TO DISEASE PROGRESSION. IT WAS REPORTED THAT A RECENT CT DEMONSTRATED THAT THE STENT GRAFT HAS MIGRATED 15 MM DISTALLY WITH A PROXIMAL TYPE 1 ENDOLEAK. THE PHYSICIAN IMPLANTED A TALENT AORTIC CUFF AND SUCCESSFULLY RESOLVED THE ENDOLEAK AND MIGRATION. NO ADD'L CLINICAL SEQUELAE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX STENT GRAFT WITH XCELERANT DELIVERY SYSTEM MIH MEDTRONIC CARDIOVASCULAR NA 779944

Patients

Seq Age Sex Outcome Treatment
1 71 YR Required Intervention