CVUE MULTIFOCAL CONTACT LENS
Report
- Report Number
- 1034196-2011-00001
- Event Type
- Other
- Date Received
- March 16, 2011
- Date of Event
- February 19, 2011
- Report Date
- March 15, 2011
- Manufacturer
- UNILENS CORP., USA
- Product Code
- HPX
- PMA / PMN Number
- K960926
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
NO DEVICE WAS RETURNED FOR EVALUATION. THE ACCOUNT INDICATED THAT THE LOT NUMBER OF THE DISPENSED PRODUCT WAS NOT RECORDED. REVIEW OF ACCOUNT AND PATIENT RECORDS ISOLATED FOUR POSSIBLE LOTS THAT THE PRODUCT COULD BE FROM R86014911, R86021108, R06029906, AND R06009015. OF THE FOUR LOTS, TWO LOTS (R06029906 AND R06009015) HAVE AVAILABLE INVENTORY THAT WAS EXAMINED FOR PACKAGE INTEGRITY AND STERILITY. STERILITY TESTING ON SAMPLES OF THESE TWO LOTS WILL BE CONCLUDED THE WEEK OF (B)(4) 2011. THE CONTRACT MANUFACTURER OF THE PRODUCT HAS BEEN CONTACTED AND IS PERFORMING LOT HISTORY REVIEWS OF THE PROCESSING OF THE FOUR SUSPECT LOTS FOR STERILIZATION EFFECTIVENESS AND OTHER REPORTS OF INFECTION.
THE PATIENT IS A CONTACT LENS USER, AND FELT THERE WAS SOMETHING ON THE LENS WHEN FIRST USED OUT OF THE PACKAGE, BUT USED THE LENS ANYWAY DUE TO LACK OF TIME. AFTER ONE HOUR, THE PATIENT HAD A SIGNIFICANT FOREIGN BODY SENSATION WITH BLURRY VISION, AND REMOVED THE LENS WITHIN 3 HOURS. ON (B)(6) PATIENT PRESENTED TO THE EYECARE PRACTITIONER WITH AN INFECTION IN THE RIGHT EYE, AND PRESENTED TO AN OPHTHALMOLOGIST ON (B)(6). THE INFECTION WAS DESCRIBED AS POSSIBLY (B)(6); A CULTURE TAKEN WAS (B)(6). THE TREATMENT WAS A PROPHYLACTIC ANTIBIOTIC AND BANDAGE CONTACT LENS, ADDING A CYCLOPLEGIC AGENT FOR COMFORT WITH ORAL ANTIVIRAL AGENT AND ORAL PAINKILLER. THE VISUAL PROGNOSIS IS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CVUE MULTIFOCAL CONTACT LENS | SOFT CONTACT LENS | HPX | UNILENS CORP., USA | CVUE MULTIFOCAL | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other | PATIENT USES OPTIFREE CONTACT LENS SOLUTION FOR| CLEANING AND CONDITIONING |