FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2031036 · Received March 4, 2011

Report

Report Number
1720753-2011-01961
Event Type
Malfunction
Date Received
March 4, 2011
Date of Event
February 10, 2011
Report Date
March 4, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MANUFACTURER'S SERVICE REPRESENTATIVE PERFORMED AN ON-SITE INVESTIGATION. THE GENERATOR INTERFACE BOARD BATTERY WAS REPLACED AND THE SOFTWARE WAS RELOADED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9800 SYSTEM WOULD NOT STOP FLUOROING. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1