9600
MDR report key: 2031018
·
Received March 4, 2011
Report
- Report Number
- 1720753-2011-01954
- Event Type
- Malfunction
- Date Received
- March 4, 2011
- Date of Event
- October 21, 2010
- Report Date
- March 4, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE BATTERY PACK, CHARGER BOARD, AND INTERCONNECT CABLE WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 9600 SYSTEM HAD A CHARGER ERROR AND LINES IN THE IMAGES. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9600 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |