FDA Adverse Event Malfunction Summary report: N

ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT

MDR report key: 20310121 · Received September 26, 2024

Report

Report Number
3005094123-2024-00499
Event Type
Malfunction
Date Received
September 26, 2024
Date of Event
September 10, 2024
Report Date
November 3, 2024
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
MMI
UDI-DI
00380740164157
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT INVESTIGATION FOR FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULT INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, DEVICE HISTORY RECORD REVIEW, FIELD DATA REVIEW, AND LABELING REVIEW. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. THE DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. A TICKET SEARCH BY LOT INDICATES THAT THE REAGENT LOT IS PERFORMING AS EXPECTED FOR THIS PRODUCT. A REVIEW OF THE COMPLAINT TRENDING REPORT DID NOT IDENTIFY ANY TRENDS FOR THE ISSUE FOR THE PRODUCT. A REVIEW OF THE DEVICE HISTORY RECORD DID NOT IDENTIFY ANY NON-CONFORMANCES OR DEVIATIONS WITH THE LOT NUMBER 64553UD01 AND COMPLAINT ISSUE. THE OVERALL PERFORMANCE OF THE ALINITY I STAT HIGH SENSITIVE TROPONIN-I ASSAY WAS REVIEWED USING FIELD DATA FROM CUSTOMERS WORLDWIDE. THE MEDIAN PATIENT RESULT FOR LOT 64553UD01 IS WITHIN THE ESTABLISHED LIMITS AND COMPARABLE WITH HISTORICAL LOTS IN THE FIELD. A REVIEW OF LABELING WAS PERFORMED AND FOUND TO SUFFICIENTLY ADDRESS THE CUSTOMER'S ISSUE. BASED ON THIS INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WITH THE ALINITY I STAT HIGH SENSITIVE TROPONIN-I, REAGENT LOT 64553UD01 WAS IDENTIFIED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, ALINITY I STAT HIGH SENSITIVE TROPONIN-I, LIST NUMBER 08P13-77, THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 04Z21. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULT GENERATED ON THE ALINITY I PROCESSING MODULE FOR A 10-DAY-OLD PATIENT. IT WAS NOTED THAT THE PATIENT AND THE PATIENT¿S MOTHER DO NOT HAVE ANY HEART-RELATED DISEASES AND BOTH TESTED NEGATIVE FOR AUTOANTIBODIES. THE CUSTOMER PERFORMED A DOUBLE DILUTION AND 25% PEG PRECIPITATION WHICH GENERATED NORMAL RESULT. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S REFERENCE RANGE: <241 PG/ML. ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULT = 6000 PG/ML. DOUBLE DILUTION = SHOWED THE RESULTS WERE NOT LINEAR. 25% PEG PRECIPITATION: RESULT = 60 PG/ML. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULT GENERATED ON THE ALINITY I PROCESSING MODULE FOR A 10-DAY-OLD PATIENT. IT WAS NOTED THAT THE PATIENT AND THE PATIENT¿S MOTHER DO NOT HAVE ANY HEART-RELATED DISEASES AND BOTH TESTED NEGATIVE FOR AUTOANTIBODIES. THE CUSTOMER PERFORMED A DOUBLE DILUTION AND 25% PEG PRECIPITATION WHICH GENERATED NORMAL RESULT. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S REFERENCE RANGE: <241 PG/ML. ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULT = 6000 PG/ML. DOUBLE DILUTION = SHOWED THE RESULTS WERE NOT LINEAR. 25% PEG PRECIPITATION: RESULT = 60 PG/ML. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1166264 ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ABBOTT IRELAND DIAGNOSTICS DIVISION 64553UD01 00380740164157

Patients

Seq Age Sex Outcome Treatment
1 10 DA Unknown ALNTY I PROCESSING MODU, 03R65-01, (B)(6)| ALNTY I PROCESSING MODU, 03R65-01, (B)(6)