TERUMO TCM II COOLING AND HEATING SYSTEM

MDR report key: 2030832 · Received March 4, 2011

Report

Report Number
1828100-2011-00468
Event Type
Malfunction
Date Received
March 4, 2011
Date of Event
February 8, 2011
Report Date
March 4, 2011
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
K883603
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

DURING PREPARATION OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE USER REPORTED THAT THE WATER FLOW FROM THE UNIT WAS VERY LOW. THE BIOMED DESCALED THE UNIT A COUPLE OF TIMES, CLEANED THE FILTERS AND SCREENS. THE WATER FLOW WAS STILL LOW IN ALL MODES OF OPERATION. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED FOR THE PROCEDURE. THE USER REPORTED THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WERE NO ADVERSE CONSEQUENCES TO THE PT AS A RESULT OF THIS EVENT.