FDA Adverse Event Malfunction Summary report: N

IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM

MDR report key: 20308291 · Received September 26, 2024

Report

Report Number
0002023141-2024-03068
Event Type
Malfunction
Date Received
September 26, 2024
Report Date
February 2, 2025
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019515
PMA / PMN Number
K061410
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT B5: DESCRIBE EVENT OR PROBLEM G3: DATE RECEIVED BY MANUFACTURER G6: TYPE OF REPORT H1: TYPE OF REPORTABLE EVENT H2: FOLLOW UP TYPE H3: DEVICE EVALUATED BY MANUFACTURER H6: ADVERSE EVENT PROBLEM H10: ADDITIONAL NARRATIVE ZIMVIE RECEIVED ONE (1) TSVB11, (IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM) FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICE IDENTIFIED THE IMPLANT OUTER PACKAGING WAS ALREADY OPENED. THE IMPLANT OUTER VIAL HAS A CRACKED GREEN CAP, AND THE OUTER VIAL TAMPER SEAL WAS INTACT. THE COOB IS CONFIRMED. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. THIS COMPLAINT REFERS TO THE SPECIFIC DEVICE BEING INVESTIGATED FOR THIS COMPLAINT RECORD. DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1260333. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1260333 FOR SIMILAR EVENTS AND ONE (1) OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: PACKAGING DAMAGED. THE CUSTOMER DID SUBMIT FOUR (4) IMAGES FOR THE REPORTED EVENT. BASED ON THE INVESTIGATION AND ACCORDING TO THE CAPA, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS EXTERNAL FACTORS (TEMPERATURE). THIS IS AN ONGOING ISSUE WHICH IS CURRENTLY BEING MONITORED. NON CONFORMANCE AND HEALTH HAZARD EVALUATION HAVE BEEN OPENED TO FURTHER CONTAIN AFFECTED PRODUCTS, EVALUATE POTENTIAL ROOT CAUSES, AND CONSIDER APPLICABLE CORRECTIVE ACTIONS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A PACKAGING MALFUNCTION DID OCCUR. THE IMPLANT'S GREEN CAP WAS CRACKED WHILE THE GREEN CAPS TAMPER SEAL WAS INTACT. THE REPORTED EVENT/COOB WAS CONFIRMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER CMP (B)(4). B3: DATE OF EVENT UNKNOWN / NOT PROVIDED. D10. CONCOMITANT MEDICAL PRODUCTS. TSVB10, IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM, LOT 1261974. TSVB11, IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM LOT 1264279. TSX31B8, TSX IMPLANT, 3.1MMD, 8MML LOT 1259848. TSX31B8, TSX IMPLANT, 3.1MMD, 8MML LOT 1261974. TSV4B10, IMP,TSV,4.1MM,SBM,10 LOT 1261280. TSV4B11, IMP,TSV,4.1MM,SBM,11.5 LOT 1262306. TSV4B13, IMP,TSV,4.1MM,SBM,13 LOT 1260099. TSVB11, IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM LOT 1264279 X 4. TSVB11, IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM LOT 1260333 X 1. TSV4B11, IMP, TSV,4.1MM, SBM,11.5 LOT 1262306 X 2. G4: PREMARKET IDENTIFICATION K013227. D10. THERAPY DATE UNKNOWN.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

DOCTOR REPORTED THAT IN THE LAST ORDER THEY RECEIVED THERE WERE SOME IMPLANTS WITH A CRACKED CAP. DOCTOR NOTICED AT THE TIME OF SURGERY AND IMPLANTS WERE CHANGED BY DISTRIBUTOR FOR NEW ONES. NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2239671 IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM DENTAL IMPLANT DZE ZIMMER DENTAL 1260333 00889024019515

Patients

Seq Age Sex Outcome Treatment
1 NA Prefer Not To Disclose DENTAL IMPLANT SEE H10 NARRATIVE