FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2030745 · Received March 22, 2011

Report

Report Number
2531779-2011-01815
Event Type
Malfunction
Date Received
March 22, 2011
Report Date
February 23, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS TORN BETWEEN THE DOWN AND OK KEYPAD BUTTONS. ALL KEYPAD BUTTONS REQUIRED EXCESSIVE FORCE TO ACTIVATE A RESPONSE DURING TESTING. ALL KEYPAD BUTTONS DO NOT SPRING BACK WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD IS UNRESPONSIVE. THE PATIENT REPORTED THAT THE KEYPAD IS NOW TORN FROM HAVING TO CONTINUALLY PUSH BUTTONS TO GET THEM TO RESPOND. IT WAS REPORTED THAT THE PUMP IS NOT EXPOSED TO WATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1