FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2030745
·
Received March 22, 2011
Report
- Report Number
- 2531779-2011-01815
- Event Type
- Malfunction
- Date Received
- March 22, 2011
- Report Date
- February 23, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS TORN BETWEEN THE DOWN AND OK KEYPAD BUTTONS. ALL KEYPAD BUTTONS REQUIRED EXCESSIVE FORCE TO ACTIVATE A RESPONSE DURING TESTING. ALL KEYPAD BUTTONS DO NOT SPRING BACK WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD IS UNRESPONSIVE. THE PATIENT REPORTED THAT THE KEYPAD IS NOW TORN FROM HAVING TO CONTINUALLY PUSH BUTTONS TO GET THEM TO RESPOND. IT WAS REPORTED THAT THE PUMP IS NOT EXPOSED TO WATER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |