FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2030733
·
Received March 22, 2011
Report
- Report Number
- 2531779-2011-01775
- Event Type
- Malfunction
- Date Received
- March 22, 2011
- Report Date
- February 22, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. DURING EVALUATION, THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION.
Description of Event or Problem · 1
PUMP WAS RETURNED TO ANIMAS FOR A SHORT LOAD STEP CAUSING A LARGE PRIME VOLUME. EVALUATION REVEALED AN OUT OF CALIBRATION FORCE SENSOR. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |