ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
Report
- Report Number
- 2531779-2011-01786
- Event Type
- Malfunction
- Date Received
- March 22, 2011
- Date of Event
- January 1, 2011
- Report Date
- February 22, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. ALL KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE DURING TESTING. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS SPRING BACK AS EXPECTED WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS.
IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PT REPORTED THAT THE UP ARROW, DOWN ARROW, AND OK KEYPAD BUTTONS HAVE BECOME INTERMITTENTLY UNRESPONSIVE OVER THE PAST SEVERAL WEEKS. HE NOTED THAT THE BUTTONS INTERMITTENTLY CLICK AND SPRING BACK AS EXPECTED. THE PT DENIED ANY PHYSICAL DAMAGE TO THE RUBBER KEYPAD; DENIED EXPOSING THE PUMP TO MOISTURE; AND STATED THAT HE WEARS THE PUMP IN HIS POCKET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR |