FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2030712 · Received March 22, 2011

Report

Report Number
2531779-2011-01786
Event Type
Malfunction
Date Received
March 22, 2011
Date of Event
January 1, 2011
Report Date
February 22, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. ALL KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE DURING TESTING. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS SPRING BACK AS EXPECTED WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PT REPORTED THAT THE UP ARROW, DOWN ARROW, AND OK KEYPAD BUTTONS HAVE BECOME INTERMITTENTLY UNRESPONSIVE OVER THE PAST SEVERAL WEEKS. HE NOTED THAT THE BUTTONS INTERMITTENTLY CLICK AND SPRING BACK AS EXPECTED. THE PT DENIED ANY PHYSICAL DAMAGE TO THE RUBBER KEYPAD; DENIED EXPOSING THE PUMP TO MOISTURE; AND STATED THAT HE WEARS THE PUMP IN HIS POCKET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 24 YR