FDA Adverse Event Malfunction Summary report: N

TECNIS IOL

MDR report key: 20307024 · Received September 25, 2024

Report

Report Number
3012236936-2024-000262
Event Type
Malfunction
Date Received
September 25, 2024
Date of Event
August 8, 2024
Report Date
December 24, 2024
Manufacturer
AMO PUERTO RICO MFG. INC.
Product Code
HQL
UDI-DI
05050474531826
PMA / PMN Number
P980040
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CORRECTED DATA: UPON FURTHER REVIEW OF THE REPORTED ISSUE, THE SUSPECT PRODUCT WAS UPDATED AND THE LENS MODEL AR40E WAS CHANGED TO LENS MODEL ZCB00, SN (B)(6) WHICH IS THE SAME IOL THAT WAS PREVIOUSLY REPORTED UNDER THE MANUFACTURER REPORT NUMBER 3012236936-2024-0001390 (J&J COMPLAINT FOLDER # (B)(4)). THIS SUPPLEMENTAL MDR IS BEING SUBMITTED AS A CORRECTION AND THE FOLLOWING FIELDS WERE UPDATED ACCORDINGLY: SECTION D1: BRAND NAME: TECNIS IOL. SECTION D4: MODEL NUMBER: ZCB00. SECTION D4: CATALOG NUMBER: ZCB0000255. SECTION D4: SERIAL NUMBER: (B)(6). SECTION D4: EXPIRATION DATE: AUG 5, 2027. SECTION D4: UDI NUMBER: (B)(4). SECTION H4: DEVICE MANUFACTURE DATE: AUG 5, 2023. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: AFTER FURTHER REVIEW, IT WAS IDENTIFIED THAT THE ZCB00 BOX WHICH THE CUSTOMER RECEIVED WAS THE SAME PRODUCT THAT HAD BEEN RETURNED BY ANOTHER CUSTOMER. SECTION A2, A3B, A4, A5 AND A6: NOT APPLICABLE, AS THERE WAS NO PATIENT CONTACT WITH THE DEVICE. SECTION D6A: IF IMPLANTED, GIVE DATE: NOT APPLICABLE, THE LENS WAS NOT IMPLANTED. SECTION D6B: IF EXPLANTED, GIVE DATE: NOT APPLICABLE, THE LENS WAS NOT IMPLANTED. HENCE, NOT EXPLANTED. SECTION E1: PHONE NUMBER: (B)(6). ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC., HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: SECTION D9: DEVICE AVAILABLE FOR EVALUATION: YES. SECTION D9: RETURNED TO MANUFACTURER ON: OCT 14, 2024. SECTION H3: DEVICE EVALUATED BY MANUFACTURER: YES. DEVICE EVALUATION: A CUSTOMER PHOTO WAS PROVIDED IN THE CP FILE AND EVALUATED. THE PHOTO DISPLAYS THE DAISY WHEEL OF AN AR40E LENS (SN: (B)(6) ON TOP OF THE ORIGINAL FOLDING CARTON FOR A ZCB00 LENS (SN: (B)(6). A 1PC LENS CAN BE OBSERVED OUTSIDE OF THE DAISY WHEEL. A 1-PIECE (1PC) LENS WAS RECEIVED INSIDE AN AR40E LENS DAISY WHEEL. PATIENT STICKERS, IMPLANT NOTIFICATION CARD, DFU, PATIENT IMPLANT CARD, AND ORIGINAL FOLDING CARTON FOR THE 1-PIECE LENS WAS ALSO RECEIVED. VISUAL INSPECTION OF THE LENS REVEAL THAT IT WAS RECEIVED DAMAGED AND WITH ONE HAPTIC DAMAGED. THE LENS WAS CLEANED AND INSPECTED; A SCRATCH WAS IDENTIFIED. THE COMPLAINT ISSUE "STERILITY ASSURANCE CONCERNS" WAS IDENTIFIED DURING PRODUCT EVALUATION; HOWEVER, BASED ON THE COMPLAINT INVESTIGATION RESULTS THE COMPLAINT ISSUE COULD NOT BE CONFIRMED TO BE RELATED TO THE MANUFACTURING OR DESIGN PROCESS AS THE CUSTOMER SENT THE PRODUCT BACK TO 3PL (3RD PARTY LOGISTICS). THEREFORE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY OR PRODUCT MALFUNCTION. THE SHIPPING HISTORY FOR THE 1-PIECE (ZCB00; SN: (B)(6) LENS WAS REVIEWED REVEALING THE LENS WAS DELIVERED TO ACCOUNT (B)(4) ON (B)(6) 2024. THE PRODUCT WAS THEN DELIVERED TO THE COMPLAINT CUSTOMER (ACCOUNT (B)(4) ON (B)(6) 2024. THE SHIPPING HISTORY OF THE 3-PIECE (AR40E; SN: (B)(6) LENS WAS THEN REVIEWED REVEALING THAT THE LENS WAS DELIVERED TO ACCOUNT (B)(4) ON (B)(6) 2023. BASED ON THE SHIPPING HISTORY BOTH 1-PIECE (ZCB00) LENS AND 3-PIECE (AR40E) LENS WERE DELIVERED TO CUSTOMER ACCOUNT (B)(4) (THIS ACCOUNT SUBSEQUENTLY RETURNED THE 1PC LENS TO J&J 3RD PARTY LOGISTICS). BASED ON THIS REVIEW, THE MIX UP OF THE LENSES MAY HAVE OCCURRED WHILE ACCOUNT (B)(4) WAS IN POSSESSION OF BOTH LENSES. ADDITIONAL INFORMATION FROM THE COMPLAINT FILE REVEALED, THE 1-PIECE LENS WAS RETURNED BY ACCOUNT (B)(4) IN THE ORIGINAL 1PC FOLDING CARTON BUT INSIDE THE ACRYLIC OF THE AR40E LENS. WHEN RECEIVING THE RETURNED PRODUCT, THE 3RD PARTY LOGISTICS DID NOT NOTICE THE COMPLAINT AND THE LENS WAS REINCORPORATED IN THE STOCK OF PRODUCT FOR SALES ON JUNE 13, 2024. THE 1PC LENS WAS SOLD TO THIS COMPLAINT¿S CUSTOMER ON (B)(6) 2024 (ACCOUNT. (B)(4). A NONCONFORMANCE (NR-0232667) WAS RAISED TO INVESTIGATE THE MISTAKE IN THE PROCESS FOR COMPLAINT PRODUCT RETURNS MANAGED BY 3RD PARTY LOGISTICS WITH J&J CUSTOMER SERVICE OVERSIGHT. AS A RESULT OF THE INVESTIGATION, ACTIONS HAVE BEEN IDENTIFIED AND ALL ACTIONS ARE IN THE PROCESS OF BEING COMPLETED. MANUFACTURING RECORD REVIEW: THE MANUFACTURING RECORDS FOR THE PRODUCT WERE REVIEWED. THE PRODUCT WAS MANUFACTURED AND RELEASED ACCORDING TO SPECIFICATIONS. A SEARCH OF COMPLAINTS RELATED TO THIS PRODUCTION ORDER (PO) WAS PERFORMED. THE SEARCH REVEALED TWO ADDITIONAL COMPLAINTS; ONE COMPLAINT IS NOT RELATED TO THIS ISSUE. ANOTHER COMPLAINT: (B)(4) FOR "HAPTIC DAMAGED, LENS DAMAGED " WHICH IS THE DUPLICATE SN TO THIS CURRENT COMPLAINT; HOWEVER, FROM A DIFFERENT CUSTOMER WAS RECEIVED ON THIS SERIAL NUMBER WHICH TWO SEPARATE INVESTIGATIONS WERE COMPLETED. NO FURTHER ACTIONS ARE REQUIRED. A SEPARATE MDR REPORT WAS ALREADY SUBMITTED FOR THE REPORTED EVENT OF "HAPTIC DAMAGED, LENS DAMAGED " WHICH WAS CAPTURED IN COMPLAINT FILE (B)(4) UNDER THE MANUFACTURER REPORT NUMBER#: 3012236936-2024-0001390. CONCLUSION: BASED ON THE INVESTIGATION, NO MALFUNCTION OR PRODUCT DEFICIENCY WAS IDENTIFIED. CORRECTED DATA: IN REVIEW, IT WAS NOTICED THAT THE FOLLOWING FIELD WAS INADVERTENTLY NOT ADDRESSED IN THE INITIAL OR FOLLOW UP #1 MDR REPORT. IN REVIEW, IT WAS ALSO NOTED THAT THE AN INCORRECT ESTABLISHMENT NAME (B)(6) WAS INADVERTENTLY ENTERED IN SECTION "E1" OF THE INITIAL MDR REPORT; THEREFORE, THE INFORMATION HAS BEEN CAPTURED IN THIS SUPPLEMENTAL MDR REPORT AS INDICATED BELOW: SECTION A3A: NOT APPLICABLE, AS THERE WAS NO PATIENT CONTACT WITH THE DEVICE. SECTION E1: ESTABLISHMENT NAME: (B)(6). ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC., HAS BEEN SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CUSTOMER ORDERED A ZCB00 LENS AND THE BOX ARRIVED SEALED. WHEN THE CUSTOMER OPENED THE MONOFOCAL INTRAOCULAR LENS (IOL), MODEL ZCB00, 25.5 DIOPTER BOX, NOTICED THAT THE DAISY WHEEL THAT CONTAINS THE LENS WAS LABELED AS AR40E, +25.5 DIOPTER. IT WAS CONFIRMED THAT THERE WAS NO PATIENT CONTACT WITH THE DEVICE. NO MEDICAL OR SURGICAL INTERVENTIONS REQUIRED, NO INJURY TO THE PATIENT REPORTED AS A RESULT OF THIS ISSUE. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1098178 TECNIS IOL INTRAOCULAR LENS HQL AMO PUERTO RICO MFG. INC. ZCB00 05050474531826

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown