FDA Adverse Event Malfunction Summary report: N

TOTALCARE SPO2RT

MDR report key: 2030650 · Received March 22, 2011

Report

Report Number
1824206-2011-01765
Event Type
Malfunction
Date Received
March 22, 2011
Date of Event
February 25, 2011
Report Date
February 25, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH REPLACED TWO STICKING SOLENOID VALVES ON THE HEAD UP AND THE HEAD HIGH VALVES TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE HEAD SECTION IS STUCK IN THE DOWN POSITION. THE NURSE IS USING REVERSE TRENDELENBURG TO KEEP THE PT'S HEAD ELEVATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE SPO2RT AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1900

Patients

Seq Age Sex Outcome Treatment
1 UNK