FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE SPO2RT
MDR report key: 2030650
·
Received March 22, 2011
Report
- Report Number
- 1824206-2011-01765
- Event Type
- Malfunction
- Date Received
- March 22, 2011
- Date of Event
- February 25, 2011
- Report Date
- February 25, 2011
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECH REPLACED TWO STICKING SOLENOID VALVES ON THE HEAD UP AND THE HEAD HIGH VALVES TO REPAIR THE BED.
Description of Event or Problem · 1
INFO RECEIVED INDICATES THE HEAD SECTION IS STUCK IN THE DOWN POSITION. THE NURSE IS USING REVERSE TRENDELENBURG TO KEEP THE PT'S HEAD ELEVATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE SPO2RT | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |