FDA Adverse Event
Malfunction
Summary report: N
AFFINITY BED
MDR report key: 2030649
·
Received March 22, 2011
Report
- Report Number
- 1824206-2011-01737
- Event Type
- Malfunction
- Date Received
- March 22, 2011
- Date of Event
- May 7, 2010
- Report Date
- May 7, 2010
- Manufacturer
- HILL-ROM INC.
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE ACCOUNT PUSHED THE BRAKE CASTER BACK INTO PLACE. TECH INSPECTED THE BRAKE CASTER AND RE FITTED THE BRAKE CASTER RETAINING SCREW AND TIGHTENED ALL BRAKE CASTER RETAINING SCREWS TO RESOLVE THE ISSUE. THIS MDR IS A RESULT OF CAPA ACTIVITY FOR THE RETROSPECTIVE REVIEW OF COMPLAINTS.
Description of Event or Problem · 1
ACCOUNT ALLEGED THAT THE FOOT END LEFT BRAKE CASTER HAD FALLEN OFF OF THE BED. NO PT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AFFINITY BED | BIRTHING BED | HDD | HILL-ROM INC. | 3700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |