FDA Adverse Event Malfunction Summary report: N

VIDEO-OPTIK "ENDOEYE 3D", 30°

MDR report key: 20306185 · Received September 25, 2024

Report

Report Number
9610773-2024-32746
Event Type
Malfunction
Date Received
September 25, 2024
Report Date
December 19, 2024
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
GCM
UDI-DI
04042761083492
PMA / PMN Number
K193026
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

E1 - FACILITY NAME: (B)(6). E2/E3: NO INFORMATION AVAILABLE FOR THE OCCUPATION OF THE INITIAL REPORTER OR WHETHER THEY ARE A HEALTHCARE PROFESSIONAL. THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTION WAS IDENTIFIED DURING THE DEVICE EVALUATION: R-UNIT (CHARGED COUPLED DEVICE) IS BROKEN. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: R-UNIT (CHARGED COUPLED DEVICE) IS BROKEN AND THEREFORE THE PIXELSHIFT IS INCORRECT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SUBJECT DEVICE HAD INTERFERENCE IN IMAGE. THERE WERE NO REPORTS OF PATIENT HARM.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2239614 VIDEO-OPTIK "ENDOEYE 3D", 30° RIGID VIDEO LAPAROSCOPE GCM OLYMPUS WINTER & IBE GMBH WA50082A 04042761083492

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown