FDA Adverse Event
Injury
Summary report: N
ASCENSION PIP
MDR report key: 2030570
·
Received March 18, 2011
Report
- Report Number
- 1651501-2011-00011
- Event Type
- Injury
- Date Received
- March 18, 2011
- Date of Event
- January 11, 2011
- Report Date
- March 18, 2011
- Manufacturer
- ASCENSION ORTHOPEDICS, INC.
- Product Code
- NEG
- PMA / PMN Number
- H010005
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
X-RAYS WERE SUPPLIED BUT WERE NOT CONCLUSIVE IN A CAUSE. AN EVALUATION OF THE RETURNED COMPONENTS WAS COMPLETED. THE FRACTURE OF THIS COMPONENT MAY BE DUE TO IMPROPER OBLIQUE OSTEOTOMY PREPARATION. IF ADD'L INFO IS OBTAINED, A SUPPLEMENT REPORT WILL BE FILED AS APPROPRIATE. CONCLUSIONS: HEAD-STEM FRACTURES, AS OBSERVED IN THE SIZE 20 PROXIMAL COMPONENT, ARE KNOWN TO OCCUR AS A RESULT OF SURGICALLY IMPACTING AN UNSUPPORTED HEAD AS A RESULT OF IMPROPER OBLIQUE OSTEOTOMY PREPARATION.
Description of Event or Problem · 1
THE DISTRIBUTOR REPORTED THAT A PT IN (B)(6) HAD A PROXIMAL COMPONENT FRACTURED. AN X-RAY THAT WAS SUPPLIED CONFIRMED THIS OBSERVATION. WHEN THE DEVICE WAS RETURNED AN ADD'L FRACTURED PROXIMAL COMPONENT WAS RETURNED ALONG WITH A FRACTURED DISTAL COMPONENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ASCENSION PIP | FINGER PYROLYTIC CARBON PROSTHESIS | NEG | ASCENSION ORTHOPEDICS, INC. | PIP-200-20P-WW |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |