FDA Adverse Event Injury Summary report: N

VASOVIEW HEMOPRO 2 EVH SYSTEM

MDR report key: 2030558 · Received March 18, 2011

Report

Report Number
2242352-2011-00156
Event Type
Injury
Date Received
March 18, 2011
Date of Event
February 17, 2011
Report Date
February 17, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K101274
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: A VISUAL INSPECTION REVEALED THAT THE JAWS WERE SOMEWHAT BURNT. THERE WAS SOME EVIDENCE OF BLOOD, AND THE HANDLE WAS BLOODY. RESISTANCE AND JAW FORCE MEASUREMENTS PASSED SPECIFICATIONS. A FUNCTIONAL TEST EVALUATING THE THERMAL SHUT-DOWN PERFORMANCE OF THE UNIT WAS PERFORMED ON THE RETURNED DEVICE USING A REFERENCE POWER SUPPLY AND CABLE. THE DEVICE PERFORMED ACCORDING TO SPECIFICATIONS DURING THE DEVICE ACTIVATION. BASED UPON THIS OBSERVATION, THE REPORTED COMPLAINT FOR "NOT CUTTING AS WELL" COULD NOT BE CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSP REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO 2 WAS NOT CUTTING AS WELL. THE HARVESTER DID THE FIRST BRANCH AND THERE WAS SOME BLEEDING WHICH WAS RESOLVED BY THE CAUTERY DEVICE. THE NEXT BRANCH WAS FINE, BUT ON THE THIRD BRANCH THE DEVICE HIT THE SHORT CYCLE AROUND 3-4 SECONDS. THE HARVESTER DID A COUPLE MORE BRANCHES BEFORE SWITCHING TO HEMOPRO 1 TO THE COMPLETE THE PROCEDURE. THERE WERE NO PT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VASOVIEW HEMOPRO 2 EVH SYSTEM ENDOSCOPIC VESSEL HARVESTING SYSTEM GEI MAQUET CARDIOVASCULAR, LLC VH-4000 25028331

Patients

Seq Age Sex Outcome Treatment
1 NA Other