FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2030549 · Received March 9, 2011

Report

Report Number
1720753-2011-02077
Event Type
Malfunction
Date Received
March 9, 2011
Date of Event
February 14, 2011
Report Date
March 9, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE COLLIMATOR AND FILAMENT WERE CALIBRATED AND THE MEMORY WAS RESET DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9900 SYSTEM HAD A COLLIMATOR AND FILAMENT ERROR MESSAGE DURING A CASE. THE PROCEDURE WAS COMPLETED. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1