FDA Adverse Event Malfunction Summary report: N

RESTORE PRIME ADVANCED

MDR report key: 2030441 · Received March 9, 2011

Report

Report Number
3004209178-2011-01771
Event Type
Malfunction
Date Received
March 9, 2011
Date of Event
February 23, 2011
Report Date
February 23, 2011
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LGW
PMA / PMN Number
P840001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

HIGH OUT-OF-RANGE IMPEDANCE INDICATIVE OF OPEN CIRCUITS WERE NOTED ON SOME OF THE BIPOLAR ELECTRODE PAIRS. THE IMPLANTABLE NEUROSTIMULATOR BATTERY WAS AT OR NEAR END OF SERVICE. IT WAS REPLACED (REFERENCE 6000032201009856). AFTER REPLACEMENT, POST OPERATIVE IMPEDANCES REMAINED OUT OF RANGE. THE CONTINUED PATTERN HIGH IMPEDANCES SUGGESTS THE CAUSE OF THE PROBLEM LAY WITH A LEAD OR EXTENSION. ADD'L INFO HAS BEEN REQUESTED. A F/U REPORT WILL BE SUBMITTED IF ADD'L INFO BECOMES AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESTORE PRIME ADVANCED LGW MDT PUERTO RICO OPERATIONS CO., JUNCOS 37702 NA

Patients

Seq Age Sex Outcome Treatment
1 40 YR LOT# NFE114665H| LEAD: MODEL 3487A, LOT# J0120768V| EXPLANTED:| IMPLANTED:| IMPLANTED:| EXPLANTED:| EXPLANTED:| LEAD: MODEL 3487A, LOT# J0118495V| IMPLANTED:| EXPLANTED:| IMPLANTABLE NEURO STIMULATOR: MODEL 7427| PROGRAMMER: MODEL 37743, LOT# NKE160032N| IMPLANTED:| ADAPTER: MODEL 74002, LOT# N251312| EXPLANTED:| PROGRAMMER: MODEL 7435, LOT# NFT012374P| EXPLANTED:| IMPLANTED:| EXTENSION: MODEL 7489, LOT# NHU008138V| EXTENSION: MODEL 7489, LOT# NHU008135V| EXPLANTED:| IMPLANTED:| IMPLANTED: