FDA Adverse Event
Malfunction
Summary report: N
RESTORE PRIME ADVANCED
MDR report key: 2030441
·
Received March 9, 2011
Report
- Report Number
- 3004209178-2011-01771
- Event Type
- Malfunction
- Date Received
- March 9, 2011
- Date of Event
- February 23, 2011
- Report Date
- February 23, 2011
- Manufacturer
- MDT PUERTO RICO OPERATIONS CO., JUNCOS
- Product Code
- LGW
- PMA / PMN Number
- P840001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
HIGH OUT-OF-RANGE IMPEDANCE INDICATIVE OF OPEN CIRCUITS WERE NOTED ON SOME OF THE BIPOLAR ELECTRODE PAIRS. THE IMPLANTABLE NEUROSTIMULATOR BATTERY WAS AT OR NEAR END OF SERVICE. IT WAS REPLACED (REFERENCE 6000032201009856). AFTER REPLACEMENT, POST OPERATIVE IMPEDANCES REMAINED OUT OF RANGE. THE CONTINUED PATTERN HIGH IMPEDANCES SUGGESTS THE CAUSE OF THE PROBLEM LAY WITH A LEAD OR EXTENSION. ADD'L INFO HAS BEEN REQUESTED. A F/U REPORT WILL BE SUBMITTED IF ADD'L INFO BECOMES AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RESTORE PRIME ADVANCED | LGW | MDT PUERTO RICO OPERATIONS CO., JUNCOS | 37702 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | LOT# NFE114665H| LEAD: MODEL 3487A, LOT# J0120768V| EXPLANTED:| IMPLANTED:| IMPLANTED:| EXPLANTED:| EXPLANTED:| LEAD: MODEL 3487A, LOT# J0118495V| IMPLANTED:| EXPLANTED:| IMPLANTABLE NEURO STIMULATOR: MODEL 7427| PROGRAMMER: MODEL 37743, LOT# NKE160032N| IMPLANTED:| ADAPTER: MODEL 74002, LOT# N251312| EXPLANTED:| PROGRAMMER: MODEL 7435, LOT# NFT012374P| EXPLANTED:| IMPLANTED:| EXTENSION: MODEL 7489, LOT# NHU008138V| EXTENSION: MODEL 7489, LOT# NHU008135V| EXPLANTED:| IMPLANTED:| IMPLANTED: |