FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 2030349 · Received March 7, 2011

Report

Report Number
9680959-2011-00692
Event Type
Malfunction
Date Received
March 7, 2011
Date of Event
December 20, 2010
Report Date
March 7, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE GENERATOR WAS REPLACED DURING THE SERVICE CALL.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 7900 SYSTEM HAD A GENERATOR FAULT ERROR MESSAGE. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1