Report Number
1030489-2024-01194
Event Type
Malfunction
Date Received
September 25, 2024
Date of Event
September 3, 2024
Report Date
December 18, 2024
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC
Product Code
KWQ
UDI-DI
00643169754584
PMA / PMN Number
K170679
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SECTION E: INITIAL REPORTER IS UNKNOWN. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS:PART # 6550017, LOT # K23G1394 VISUAL AND OPTICAL INSPECTION CONFIRMED THE END OF THE TAB EXTENDER THAT INTERFACES WITH THE BREAK OFF SCREW HAS BEEN BENT. THIS TYPE OF DAMAGE IS CONSISTENT WITH BEND STRESS OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING AN EXTENDER USED FOR LUMBAR PERC UTANEOUS POSTERIOR DECOMPRESSION FIXATION. IT WAS REPORTED THAT THE CONNECTION TO THE SCREW IS SLIGHTLY BENT, SO THE SCREW IS NOT ATTACHED. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2263060 CD HORIZON® SOLERA® VOYAGER¿ SPINAL SYSTEM INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED M KWQ MEDTRONIC SOFAMOR DANEK USA, INC 6550017 K23G1394 00643169754584

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown